| MICROPROCESSOR BASED INFRARED GAS ANALYZER Model ZRF (Single or Dual Component) |

| EXCLUSIVE ZIRCONIA O2 SENSOR (separate purchase required) |
MODEL ZFK3
For O2 correction, the gas analyzer Model ZRF can accept a linearized 0 to 1
VDC signal coming from any analyzer calibrated 0 to 25% O2 full scale. If an
analyzer is not available, CAI can supply the Exclusive Zirconia O2 Sensor Model
ZFK3.
MEASURABLE COMPONENT & MEASURING RANGE:
| MEASUREABLE COMPONENT | 1ST RANGE | 2ND RANGE |
| OXYGEN | 0 TO 10 Vol% | 0 TO 25 Vol% |
|
SPECIFICATIONS
MEASURING METHOD: Zirconia System REPEATABILITY: Within ± 0.5% of full scale LINEARITY: Within ± 2% of full scale ZERO DRIFT: Within ± 1% of full scale/week SPAN DRIFT: Within ± 2% of full scale/week RESPONSE TIME: Approx. 20 seconds (for 90% response) SAMPLE FLOW RATE: 0.5 ± 0.25 L/min. GAS INLET/OUTLET SIZE: NPT 1/4 POWER SUPPLY: 90 to 126VAC or 200 to 240VAC, 50/60HZ ENCLOSURE: Steel casing, for indoor application INDICATOR: Temperature indication (LED) TEMPERATURE ALARM OUTPUT: One for contact capacity 220V, 1A AC (resistive load) DIMENSIONS: 5.5" H x 6.6" W x 7.5" D (140mm H x 168mm W x 191mm D) WEIGHT: 6.6 lbs. (3 kg) Specifications subject to change without notice |