RAD7 Electronic Radon Monitor/Sniffer


Highlights

RAD7 is everything you ever wanted in an electronic radon monitor.

 
FEATURE Specification
MULTI-MODES  Continuous Radon Gas Monitor
Long-term/Short-term screener 
Sniffer, with small tube "snout", to search for radon and thoron entry points.
MULTI-MEASUREMENTS Measures radon in air, soil, and water (with RAD H2O accessory)
SENSITIVITY Monitor: 0.4 counts/min/pCi/l
Sniffer: 0.2 counts/min/pCi/l
RANGE 0.1- ----> 20,000 pCi/l
MEMORY 1,000 previous radon concentrations, and associated data. Can be read out on LCD, down loaded and/or printed out. Also shows high, low, average and standard deviation of readings.
PRINCIPLE OF OPERATION Electrostatic collection of alpha-emitters with spectral analysis.
BUILT-IN-AIR PUMP Nominal 1 liter/minute flow rate. Inlet air filter
Inlet and outlet air connectors.
SPECTRUM READOUT* Pulse height spectrum of alpha distribution verifies radon/thoron, and also shows the RAD7 is operating properly. Accurate determination of the alpha particle energies creates a radon signature that allows discrimination of radon and thoron's alpha particles from those of other isotopes.
FAST LOW-LEVEL READINGS* In Continuous Monitor mode, RAD7 measures the EPA action level of 4 pCi/l in just 1 hour (60 minutes), with standard deviation of 10%. Since the RAD7 has virtually no background, it is much more sensitive than other electronic detectors, easily measuring down to 0.1 pCi/l
RECOVERY IN MINUTES* Recovers from high radon exposure with a 3.05 minute half-life: to less than 10% of peak value in 12 minutes; to less than 1% of peak value in 30 minutes. Drops from 20,000 to 1 pCi/l in 8 hours.
TAMPER PROOF* A "Test Lock" command is programmed into the machine to secure your RAD7 against all kinds of intentional or unintentional tampering.
AUTO MODE* This setting starts a test run in Sniff mode, then switches to Normal mode after the first 3 hours. No more waiting for equilibrium. Gives quick response, followed by statistical precision.
CONTINUOUS UPDATING* Gives continuous update of radon levels so you can see the trend at all times, not just at pre-set intervals.
COMPUTER LCD DISPLAY 2 line X 16 character, alpha-numeric. easy to read
AUDIBLE RADON COUNT Indicates presence and intensity of radon and thoron. May be turned on or off.
POWER AC/Battery powered - 5 AH 6V batteries. Automatic charge when plugged in. Continuous battery operation: 24 hours in Sniffer mode; 72 hours in Monitor mode.
PRINTER Hewlett-Packard Model HP 82240A
RS-232 PORT Computer/modem hookup
OPERATING RANGE Temperature: 40° - 105° F.(5° - 40° C.)
Humidity: 0 - 100%, non-condensing.
WEIGHT/SIZE 11 pounds complete - 9.5" X 7.5" X 10.5"
CASE MATERIAL High density polyethylene.

* RAD7 exclusive feature


  back to top

 

 

 

home sitemap EAS service links testimonials

contact us

 



Environmental Analytical Systems
286 Mask Island Drive
Barry's Bay, Ontario, Canada K0J 1B0

Phone: 613 756 0101
Toll Free (Canada only): 1 800 427 8591
Fax: 613 756 0909
email: info@enviro-analytical.com