These modules are designed for incorporation in OEM instrument applications. An RS-232 interface and software interface guide is provided. An analog output may also be used with a buffer amplifier. Input power is +5, +/- 15 V.
The X1003 is designed for anesthesia gas monitors. It is accurate in any mixture of anesthetic gases and independent ambient temperature changes and flow system induced pressure changes. The 10 to 90% response time is 130 ms at a flow of 200 ml/min.
The X1004 is designed for metabolic gas exchange measurements. It has a resolution of 0.02% oxygen and accuracy of +- 0.2% in air mixtures with a response time of 100 ms at 200 ml/min.
The X1007 is designed for aircraft applications and can measure oxygen concentrations at sample and/or ambient pressures of 20 to 115 kPa.